Prosonic FDU9x FDU91F
Brand: Endress+HauserProsonic FDU9x FDU91F
Brand: Endress+Hauser
Extended Specs at a Glance
Max. measurement distance:
Liquids: 10 m (33 ft),
Solids: 5 m (16 ft)
Solids: 5 m (16 ft)
Process pressure / max. overpressure limit:
0.7 bar ... 4 bar abs
(10 psi ... 58 psi)
(10 psi ... 58 psi)
Main wetted parts:
316L (fully welded IP68 / NEMA 6P)
Accuracy:
+/- 2 mm + 0.17% of measured distance
Process temperature:
-40 °C ... 105 °C
(-40 °F ... 221 °F)
CIP: 30 min: 135 °C
(30 min: 275 °F)
(-40 °F ... 221 °F)
CIP: 30 min: 135 °C
(30 min: 275 °F)
Field of application
The FDU91F ultrasonic sensor for continuous, non-contact and maintenance-free level measurement of fluids, pastes, sludges and powdery to coarse bulk materials. Also for flow measurement in open channels and at weirs. Measurement is unaffected by dielectric constant, density or humidity and also unaffected by build-up due to the self-cleaning effect of sensors. Fully welded 316L with hygienic process connection available. Maximum measuring range in liquids 10m (33ft), solids 5m (16ft).
Key Benefits
- Integrated temperature sensor for Time-of-Flight correction. Accurate measurements are possible, even if temperature changes are present
- Hermetically welded sensors (316L, 1.4404, 1.4435) for highest chemical resistance
- Suited for rough ambient conditions thanks to separate installation from the transmitter
- Reduced build-up formation because of the self-cleaning effect
Specifications
Brand
Industry
- Chemical
- Food & Beverage
- Mining, Minerals & Metals
Product Family
- Prosonic
Continuous / Liquids
Characteristic / Application
- Separated version with field housing or top hat rail housing for control cabinet instrumentation, 300 m in-between sensor and transmitter
Supply / Communication
- 4-wire (HART, PROFIBUS DP)
Accuracy
- +/- 2 mm + 0.17% of measured distance
Ambient temperature
- -40 °C ... 105 °C
- (-40 °F ... 221 °F)
Process temperature
- -40 °C ... 105 °C
- (-40 °F ... 221 °F)
- CIP: 30 min: 135 °C
- (30 min: 275 °F)
Process pressure / max. overpressure limit
- 0.7 bar ... 4 bar abs
- (10 psi ... 58 psi)
Main wetted parts
- 316L (fully welded IP68 / NEMA 6P)
Process connection
- G / NPT 1", DN80 / 100, ASME 3" / 4"
Process connection hygienic
- Tri-Clamp ISO2852, DN80 3 1/2", DN100 4"
Blocking distance
- 0.3 m (1 ft)
Max. measurement distance
- Liquids: 10 m (33 ft),
- Solids: 5 m (16 ft)
Communication
- Transmitter:
- 4 ... 20 mA HART
- PROFIBUS DP
Certificates / Approvals
- ATEX, FM, CSA, IEC Ex, INMETRO, NEPSI, UK Ex
Design approvals
- EN 10204-3.1
Hygienic approvals
- 3A
Components
- Transmitter:
- FMU90, FMU95
Application limits
- Foam / high turbulence possible:
- FDU92
- For tank farms scanner:
- FMU95
Continuous / Solids
Characteristic / Application
- Separated version with field housing or top hat rail housing for control cabinet instrumentation, 300m in-between sensor and transmitter
Supply / Communication
- 4-wire (HART, PROFIBUS DP)
Accuracy
- +/- 2mm + 0.17% of measured distance
Ambient temperature
- -40 °C ... 105 °C
- (-40 °F ... 221 °F)
Process temperature
- -40 °C ... 105 °C
- (-40 °F ... 221 °F)
- 30 min: 135 °C / 275 °F)
Process pressure / max. overpressure limit
- 0.7 bar ... 4 bar abs
- (10 psi ... 58 psi)
Main wetted parts
- 316L (fully welded IP68 / NEMA 6P)
Process connection
- G / NPT 1", DN80 / 100, ASME 3" / 4"
Process connection hygienic
- Tri-Clamp ISO2852, DN80 (3 1/2"), DN100 (4")
Blocking distance
- 0.3 m (1 ft)
Max. measurement distance
- Liquids: 10 m (33 ft),
- Solids: 5 m (16 ft)
Communication
- Transmitter:
- 4 ... 20 mA HART
- PROFIBUS DP
Certificates / Approvals
- ATEX, FM, CSA, IEC Ex, INMETRO, NEPSI, UK Ex
Design approvals
- EN 10204-3.1
Options
- Second 4...20 mA output
Components
- Transmitter:
- FMU90